Trace concentration calibration standards of acetaldehyde in
carbon dioxide are dynamically blended using the FlexStream™ Automated Gas Standards Generator and TraceSource™
Permeation Tubes. Acetaldehyde emitted by the permeation tube is added
directly to a controlled flow of purified or beverage grade carbon dioxide to
form the mixture. Mixture concentration
is adjusted by varying to carbon dioxide flow. User may achieve concentrations
from over 400 ppb down to 25 ppb using a single permeation tube. Lower
concentrations, or variable concentration in constant flow, are available using
optional secondary dilution. The standard additions technique is used to
compensate for residual acetaldehyde in the carbon dioxide. Mixtures
concentrations are traceable to NIST through physical standards.
Acetaldehyde is a key component causing ‘off taste’ and ‘off
odor’ problems in carbonated beverages and is routinely monitored to ensure
concentrations below 200ppb. Trace concentration mixtures are generally not
stable in storage. The dynamic blending method using permeation tubes avoids
that problem and insures reliable standards.