World Congress on Biosensors 2014

World Congress on Biosensors 2014
Biosensors 2014

Thursday 16 September 2010

Rapid determination of DNPH derivatized

This application note describes a very fast method for identifying six DNPH derivatized
carbonyls in a mixture using the KNAUER PLATINblue UHPLC system. The stationary phase
BlueOrchid C18 A was used to separate the carbonyl mixture. Short columns with small
particles are the most suitable way for avoiding long equilibration and analysis times.
Reduction of analysis time from about 20 to less than 3 minutes compared to a
conventional HPLC method is achieved by using the BlueOrchid C18 A phase with a
1.8 μm particle size. A binary high pressure gradient configuration was used at a flow rate
of 0.8 ml/min in combination with a 2 mm column ID and UV detection

See more at: http://www.knauer.net/dwnld_fls/a_e_vev3_uhplc_dnph_carbonyls.pdf?k=21021-1

No comments:

Post a Comment