The FlexStream™ computer controlled, Gas Standards
Generating System dynamically blends calibration gas standards for trace sulfur
compounds in food grade carbon dioxide.
Trace Source™ permeation tubes add low ppb concentrations of typical
sulfur species including COS, H2S, mercaptans directly to a controlled flow of
purified, beverage grade CO2 to create trace concentration calibration
standards. Residual background
contamination in the CO2 matrix gas is
compensated for by using the Standard Additions Technique. Calibration is based on the known difference
in sulfur compound concentration due to the addition from the Trace Source™
permeation tubes. A secondary dilution
feature gives ppt concentrations.
Concentration is adjustable over a wide dynamic range. Mixture concentrations are traceable to NIST
through physical standards.
Trace sulfur impurities in carbon dioxide used for
carbonation create “off taste” and “off odor” in beverage products. Since these impurities occur naturally, the
CO2 purity must be monitored. Effective calibration of the monitoring systems
requires traceable reference standards at multiple concentrations for each of
the potential contaminants. Static
mixtures containing trace concentrations of sulfur compounds are typically not
stable in storage.
For more information, please write or call: KIN-TEK Laboratories, Inc., 504 Laurel, La Marque,
TX 77568, Phone: (409) 938-3627, Fax: (409) 938-3710,
www.kin-tek.com, email: sales@kin-tek.com