Obtaining accurate gas standards has been particularly difficult because formaldehyde vapor condenses to a solid polymer at temperatures below 80°C, resulting in unstable concentration in stored mixtures.
In Trace Source™ permeation tubes, alpha-polyoxymethylene, a high purity formaldehyde polymer, decomposes to maintain a stable vapor pressure of formaldehyde and consequently a stable emission rate of formaldehyde. Formaldehyde emitted from the Trace Source™ tube is mixed with a known flow of zero air to form the trace concentration gas standard.
Emission flow measurements are traceable to NIST through fundamental measurements (weight change, temperature, etc.). Mixing the emission flow with a traceable flow dilution gas creates a traceable mixture.
For more information, please write or call: KIN-TEK Laboratories, Inc., 504 Laurel, La Marque, TX 77568, Phone: (409) 938-3627, Fax: (409) 938-3710, www.kin-tek.com,
email: sales@kin-tek.com
No comments:
Post a Comment